Customer Requirements
PCR Thermocycler
- Steady state temp distribution using less than 15 Watts (1)
- Prefer aprox. 2 Watts
- Prefer No more than 5 watts in heating.
- Reaches steady state in less than 10 minutes (2)
- Runs off external power supply (3)
- W/o computer hookups (use microcontroller)
- Fully portable (2)
- Fit within a cubic foot (ask Neil)
- Operator interfaced with inputs for desired temperature for upper and lower limits. (1)
- 80–95 deg C
- 50 – 65 deg C
- Region of interest stays optically visible top and bottom (1)
- Wavelength =425–485
- Slides have tolerance ± 0.5 mm all directions (1)
- Slide must be interchangeable (1)
- Prefer protected from convection environment (2)
- Enclosed environment for glass slide
- Variation from linear distribution must be ± 1 deg C (1)
- Temperature across length is ± 0.5 deg C (1)
- Temperature difference from set point is ± 0.5 deg C (1)
- Middle region is optically visible (55 mm x 10 mm) (1)
- Materials must be biocompatible. (1)
- Ease of Use (“Idiot Proof”) (2)
- Upper temperature range 80 – 95 deg C
- Lower temperature range 50–65 ºC
(1) High Importance
(2) Medium Importance
(3) Low Importance
PCR Fluid Injector